의료용 초음파 팬텀을 활용한 부채꼴 탐촉자의 영상 성능 평가 : 부산지역을 중심으로

원문정보

Evaluation of Imaging Performance of Sector Transducer Using Medical Ultrasound Phantom : A Regional Study in Busan

최효은, 양성희

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

This study aimed to quantitatively evaluate the imaging performance of sector transducers used in cardiac ultrasound systems using the ATS-539 phantom and to verify the clinical applicability of phantom-based quality control (QC). Twenty-six cardiac ultrasound units were assessed for seven parameters: dead zone, vertical distance accuracy, horizontal distance accuracy, axial resolution, lateral resolution, sensitivity, and gray scale/dynamic range. All measurements were repeated five times under identical settings, and mean values were analyzed. Most systems met the recommended performance standards, whereas some older units showed reduced vertical distance accuracy and variations in gray-scale reproducibility. These findings demonstrate that phantom-based quantitative evaluation is effective for maintaining image quality in cardiac ultrasound systems and provide foundational evidence for establishing standardized QC guidelines in clinical practice.

한국어

본 연구는 심장 초음파 장비에 사용되는 부채꼴 탐촉자의 영상 성능을 ATS-539 팬텀을 활용하여 정량 적으로 평가하고, 팬텀 기반 품질관리(QC)의 임상적 적용 가능성을 확인하고자 하였다. 총 26대의 심장 초음파 장 비를 대상으로 불응영역, 수직 및 수평거리 정확도, 축방향 및 측방향 해상도, 민감도, 회색조 등 7개 항목을 동일 한 조건에서 반복 측정하여 평균값을 산출하였다. 평가 결과 대부분의 장비가 권장 기준을 충족하였으나, 일부 노 후 장비에서 수직거리 정확성 저하와 회색조 항목의 영상 재현성 차이가 확인되었다. 본 연구는 심장 초음파 장비 의 성능 편차를 정량적으로 제시함으로써, 팬텀 기반 평가의 유효성과 임상 품질관리 체계 확립의 필요성을 제시 하였다는 점에서 의의가 있다.

목차

요약
Abstract
Ⅰ. 서론
Ⅱ. 실험방법
2.1 실험재료 및 도구
2.2 ATS-539를 이용한 평가 기준 설계
2.3 영상 성능 평가 측정 항목
2.4 초음파영상 획득 및 평가
Ⅲ. 실험결과
3.1 초음파 장비의 분포 및 특성
3.2 측정 항목 평가 결과
Ⅳ. 고찰
Ⅴ. 결론
REFERENCES

저자정보

  • 최효은 Hyo-Eun Choi. 센트럴병원 영상의학과 방사선사
  • 양성희 Sung-Hee Yang. 부산가톨릭대학교 방사선학과 조교수

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.

      • 4,000원

      0개의 논문이 장바구니에 담겼습니다.