Comparison of the skill performance based on an automated external defibrillator training method : A manikin-based study
초록
영어
Purpose: The purpose of this study is to evaluate the interrupted chest compression time during the use of an automated external defibrillator (AED) depending on different AED practice training methods, and to report differences in self-efficacy before and after training. Methods: We enrolled university freshmen who have had cardiopulmonary resuscitation (CPR) training but have not or have had AED training but over 6 months. We examined differences between the group that practiced only shockable rhythms during training and the group that practiced both shockable and non-shockable rhythms. Results: A total of 72 individuals participated in this study, with 36 individuals each in the control and experimental groups. There was no statistically significant difference in the proficiency of AED usage between the two groups. In non-shockable cases, the experimental group showed shorter chest compression interruption time than the control group (2.30±1.21sec vs. 3.16±1.73 sec; p<0.01). In terms of self-efficacy before and after training, both groups showed higher self-efficacy after than before training. Conclusion: Individuals who underwent training that provided practice on both shockable and non-shockable rhythms had a shorter interrupted chest compression time when using the AED.