일회용 약물 주입기구를 이용한 5-Fluorouracil의 지속주입효과와 용기 내 안정성 평가

원문정보

Determination of Flow Rate and Stability of 5-Fluorouracil in Disposal Infusion Device, Anapa

김정태, 정성현

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

Disposal infusion device is known to be useful for chemotherapy. Anti-cancer drug can be released by the force of carbon dioxide or balloon. In this study, we compared the Anapa® (LC0020) with B Company (LV2 ml) in terms of infusion rate and stability. Infusion rate was determined every six minute using software, MSI08IH. Stability of 5-fluorouracil was examined periodically using a High Performance Liquid Chromatography. Infusion rates of gas-derived Anapa device were 2.29, 1.86, 1.98 ml/hr and those of balloon-derived B Company device were 1.71, 1.58, 1.37 ml/ min. There were no significant differences in stability of 5-fluorouracil between Anapa® and B Company devices. In summary, gas-derived Anapa® device is thought to be comparable or superior to balloon-derived B Company device as far as infusion rate and stability are concerned. We expect that Anapa® as a home infusion device can be employed to improve a quality of life and compliance of cancer patients.

목차

Abstract
 연구방법
  5-FU 주사제의 조제
  유속실험
  5-FU의 함량분석
  통계분석
 결과 및 고찰
  유속실험
  안정성 실험
 결론
 참고문헌