One of recent topics in the case of hepatitis B virus(HBV) is the value of hepatitis B surface antigen(HBsAg) concentration as a prognostic marker. We developed quantitative method of rapid immunochromatographic assay(ICA) kit for HBsAg using computer image analysis(CIA) for the purpose of home diagnosis. Quantitative ICA using CIA demonstrated integrated optical
density(IOD) values proportional to log of reference HBsAg concentrations in the range of 2∼200 ng/mL, and enzyme-linked immunosorbent assay(ELISA) demonstrated the same in the range of 0.1∼100 ng/mL, however, the test results with sample sera showed the same concentrations on both kits. Furthermore, repeated tests with the same samples revealed that this quantitative ICA using CIA would be reproducible and coefficient of variation(CV) of the results was 1.38∼6.30%.
한국어
B형간염 바이러스 연구에 관한 최근의 주제는 혈중 B형간염 표면항원농도(HBsAg)를 예후 판정의 지표로 활용하는 것이다. 본 연구에서는 B형간염에 대한 자가진단을 목적으로 하여 컴퓨터 상 분석 기법을 사용하는 HBsAg에 대한 정량적 면역크로마토그래피 측정법을 개발하였다. 본 방법을 사용하였을 때 측정된 integrated optical density(IOD)은 2∼200 ng/mL의 범위 내에서 표준 HBsAg 농도의 log값에 비례하여 증가하였데, ELISA 측정법은 0.1∼100 ng/mL의 범위 내에서 표준 HBsAg 농도의 log값에 비례하여 증가하였다. 3개의 혈청 검체를 가지고 두 가지의 측정법으로 각각 시험하여 표준곡선에 대입하였을 때 동일한 농도를 나타내었는데 그 값이 각각 9,000, 7,000, 4,000 ng/mL 이었다. 또한 컴퓨터 상 분석 기법을 사용하는 HBsAg에 대한 정량적 면역크로마토그래피 측정법으로 동일 검체를 5회 반복
시험하였을 때 측정된 IOD값이 재현성이 있었으며, 변이계수가 1.38∼6.30% 이내였다.