라돈 표준실과 가옥 내에서 일부 라돈검출기에 대한 성능 평가

원문정보

Performance Evaluation of Several Radon Detectors in the Standard Chamberand Dwellings

윤석원, 변종인, 김용재, 장병욱, 윤주용

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

To ensure the performance of radon detectors, three passive radon detectors (RadTrak®, Radopot®, and E-PERM®) have beenreviewed. The difference ratios of RadTrak and Radopot tested in the radon standard chamber were -13.2% and -6.0%, respectively,which were in good accordance within 20% of the value measured by AlphaGUARD®. To ensure the performance of the long termmeasurement, the 3 detectors were installed at the same position of approximately one hundred of dwellings for one year. Thecorrelation curve between RadTrak and Radopot shows good agreement with a correlation coefficient (R2) of 0.91. However, Thecorrelation curve between E-PERM and Radopot shows bad agreement (R2= 0.021). In addition, the distribution map of annual meanindoor gamma dose rate measured with E-PERM was not in accordance with the distribution map of outdoor gamma dose ratemeasured by Portable Ion Chamber. According to the results, some requisites for the selection of the radon passive detectors in thelarge-scale indoor radon survey were discussed.

목차

초록
 1. 서론
 2. 재료 및 방법
 3. 결과 및 논의
  3.1 측정결과
  3.2 논의
 4. 결론
 참고문헌
 abstract

저자정보

  • 윤석원 Seok-Won Yoon. 과학기술연합대학원대학교(Department of Measurement and Detection of Radiation, University of Science and Technology)
  • 김용재 Yong-Jae Kim. 과학기술연합대학원대학교, 한국원자력안전기술원(Korea Institute of Nuclear Safety)
  • 장병욱 Byung-Uck Chang. 과학기술연합대학원대학교, 한국원자력안전기술원(Korea Institute of Nuclear Safety)
  • 변종인 Jong-In Byun. 과학기술연합대학원대학교(Department of Measurement and Detection of Radiation, University of Science and Technology)
  • 윤주용 Ju-Yong Yun. 과학기술연합대학원대학교, 한국원자력안전기술원(Korea Institute of Nuclear Safety)

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      0개의 논문이 장바구니에 담겼습니다.