인공 고관절 비구컵의 끼워맞춤에 따른 변형량의 예측

원문정보

FE simulation of Deformation of Total Hip Arthroplasty Acetabular Cup Under Press-fit

안영명, 박영배

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

In a total hip arthroplasty, the artificial hip joint is composed of an acetabular cup and a femoral head. To minimize the wear of the joint, the bearing surface should be precisely spherical. There were concerns that the press-fitting of the acetabular cup to the pelvis may cause the deformation of the cup and accelerate the wear of the joint, but its in-vivo measurement was challenging. In this paper, 3 dimensional finite element(FE) models of a pelvis and acetabular cups of Metasul 50mm and Pinnacle 50-60mm cups were used to simulate the deformation of the acetabular cups. For Metasul cups, the change of inner radius with respect to the location and the maximum shrinkage of the inner radius were found. For the Pinnacle cups, maximum change
of the outer diameter were found and compared with the literature. FE model showed that the maximum shrinkage of the inner radius of the Metasul cup was 23μm (1.0mm press-fit, Bone stiffness 17GPa case).
The shrinkage occurred mainly on the anterior and posterior side of rim of the cup, and the amount was proportional to the press-fit amount. The diametric change of the Pinnacle cup was 0.16mm on average, which was in same range of the clinically reported value. In conclusion, under the normal condition the reduction of the inner radius of the Metasul cup was too small to cause the jamming or the excessive wear.

목차

Abstract
 1. 서론
 2. 대상 및 방법
 3. 해석결과
 4. 고찰
 5. 결론
 References

저자정보

  • 안영명 Y. M. AHN. 서일대학 자동차과 부교수
  • 박영배 Y. B. PARK. 큐렉소(주) 부설연구소 책임연구원