Applicability of Using GC-PDD (Pulsed Discharge Detector) for Multiresidual Pesticides Analysis

원문정보

Oh, Chang-Hwan

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

The electron-capture detector (ECD) of gas chromatographs (GC) has been used widely in pesticide analysis. However, as ECD relies on radioactive material, it is troublesome to purchase and maintain. Therefore, potent replacements for ECD were investigated. A Pulsed-discharge detector (PDD) for ECD was tested and the analytical results of PDD (ECD mode),u?CD , and nitrogen-phosphorus detector (NPD) were compared for 107 pesticides including organochroline, organophosphorus, pyrethroids etc. The number of pesticides identified at the lowest limit of detection (LOD) was 36, 29, and 2 for PDD,u?CD , and NPD, respectively. The remaining pesticides showed same response to PDD and u?CD. The GC-PDD analysis of pesticides spiked into representative agricultural products (brown rice, spinach, and mandarin oranges) also showed good and/or equivalent recoveries using GC-u?CD.

목차

Abstract
 Introduction
 Materials and Methods
 Results and Discussion
 Acknowledgments
 References