Uroflowmetry is a non-invasive clinical test useful for screening benign prostatic hyperplasia(BPH) common in the aged men. The standard way to obtain the urinary flow rate is to continuously measure the urine weight proportional to volume over time. The present study proposed an alternative technique measuring pressure to overcome noise problems present in the standard weight measurement.
Experiments were performed to simultaneously acquire both weight and pressure changes during
urination of 9 normal men. Noise components were separated from volume signals converted from both weight and pressure signals. Signal-to-noise ratio was defined as the ratio of the signal to noise power in volume changes, which was 8.5 times larger in the pressure measuring technique, implying that cleaner signal could be obtained, more immune to noisy environments. When four important diagnostic parameters were estimated, excellent correlation coefficients higher than 0.99 were resulted with mean relative errors less than 5%. Therefore, the present pressure measurement seemed valid as an alternative technique for uroflowmetry.
한국어
요류검사는 배뇨시 요의 유량을 비침습적으로 관찰하는 검사로 고령 남성의 대표적 질환인 전립선 비대증의 진단 및 관리에 유용하게 사용된다. 기존의 요류검사는 배뇨중 요(소변)의 무게를 로드셀(load cell)로 측정하여 요류신호를 산출하는데, 요 수집용기 바닥에서 무게를 측정하기 때문에 충격에 민감한 단점을 가지고 있다. 본 연구에서는 이를 극복하기 위해 수압 측정방식의 새로운 요류 계측방법을 제안하였으며, 수압 측정 방식의 요류검사로 산출한 진단매개변수를 기존의 무게 측정방식과 비교함으로써 정확도를 평가하고자 하였다. 건강한 성인 남성 9명을 대상으로 무게 및 수압 측정방식의 요류검사를 동시에 수행하여 용적신호를 산출한 후 잡음성분을 분리하였다. 두 용적신호에 대한 잡음성분의 비(signal-to-noise ratio)를 산출한 결과 수압 측정방식이 8.7배 더 크게 나타났다. 피검자별 요류신호로부터 산출한 4개의 진단 매개변수들을 무게 측정방식과 수압 측정방식 간에 선형 회귀분석하여 비교한 결과 모두 0.99 이상의 높은 상관관계를 나타내었다. 따라서 수압 측정방식의 새로운 요류검사 방법으로 정확한 진단 매개변수를 산출할 수 있음을 입증하였다.
목차
Abstract
I. 서론
II. 재료 및 방법
III. 결과