Quantitative Image Evaluation Using ATS-539 Multi-Purpose Phantom and Abdominal Phantom in Ultrasound Imaging
초록
영어
This study applied the MTF and NPS to quantitatively evaluate ultrasound image quality in terms of resolution and noise. Ultrasound images were obtained using the ATS-539 multipurpose phantom and the ABDFAN human abdominal phantom, varying the frequencies for each probe. The MTF results indicated that the spatial frequency values at 50% MTF were high in the order of frequencies Res, Gen, and Pen, especially in the ROI at the focal position. Comparing the ROIs set at the same location, the values increased as the frequency of the image increased. The NPS results showed that NPS values were high in the order of frequencies Res, Gen, and Pen for all probes. Further research is needed to establish specific MTF and NPS measurement methods and reference values by applying various ultrasound equipment and parameters.
한국어
본 연구에서는 초음파 영상 품질의 정량적인 평가를 위해 변조전달함수(MTF)와 잡음력스펙트럼(NPS)을 적용하여 해상도 및 잡음을 평가하였다. ATS-539 다목적 초음파 팬텀과 ABDFAN 인체형 복부 팬텀을 이용하 여 각각의 탐촉자별 주파수 변화에 따라 초음파 영상을 획득하였다. MTF 측정 결과 50% MTF에 해당하는 공간 주파수 값은 각 탐촉자 모두 주파수 Res, Gen, Pen 순서로 높게 나타났다. 그리고 초점 위치의 ROI에서는 공간 주파수가 높게 나타났으며, 같은 위치에 설정한 ROI별로 비교했을 때 영상의 주파수가 높아질수록 공간주파수가 높게 나타났다. NPS 측정 결과 각 탐촉자 모두 주파수 Res, Gen, Pen 순서로 NPS 값이 높게 나타났다. 초음파 영상에서 구체적인 MTF, NPS 측정 방법과 기준값을 수립하기 위해 향후 다양한 초음파 장비와 파라미터 조건 을 적용하여 활발한 연구가 이루어져야 할 것으로 사료된다.