earticle

논문검색

<학술연구>

전자동 혈액 점도계용으로 새롭게 개발한 소형 일회용 U-튜브의 분석적 성능 평가 : 검증 및 기존 방식과의 비교

원문정보

Analytical Performance Evaluation of a Newly Developed Miniaturized Disposable U-tube for Automated Blood Viscometer : Validation and Comparison to Conventional Methods

안효진, 정진무

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

This study evaluates the analytical performance of a newly developed miniaturized disposable U-tube for an automated blood viscometer and compares it to conventional viscometers. Whole blood viscosity (WBV), essential for circulatory function, exhibits non-Newtonian behavior, posing challenges for measurement at low shear rates. The blood viscometer, based on a scanning capillary tube method, used disposable U-tubes to measure viscosities across a shear rate range of 1s⁻¹ to 1,000s⁻¹. Precision evaluation showed stable coefficients of variation (CV) across different viscosity levels. Repeatability assessment indicated consistent CV values, demonstrating the reliability of the device. The agreement with the LV-III Brookfield viscometer and MCR 92 Rheometer was analyzed using Bland-Altman plots, which revealed minor systematic biases and consistent differences across the measurement range. Correlation analysis using Passing-Bablok regression showed high correlation coefficients (R > 0.96) with regression slopes close to 1. The newly developed miniaturized disposable U-tube exhibits excellent precision, reliable repeatability, and high correlation with established methods, enhancing laboratory productivity and offering potential for clinical applications. Further studies with human blood samples are recommended to confirm its clinical applicability.

목차

ABSTRACT
1. Introduction
2. Materials and Methods
2.1 Precision Evaluation
2.2 Repeatability Evaluation
2.3 Correlation Evaluation
2.4 Statistical Analysis
3. Results
3.1 Precision Evaluation
3.2 Repeatability
3.3 Correlation
4. Discussion
5. Conclusion
Declaration of Interests
Acknowledgement
References

저자정보

  • 안효진 Hyo-Jin Ahn. Corporate Research Institute, Biorheologics Co., Ltd., South Korea
  • 정진무 Jin-Mu Jung. Department of Mechanical Design Engineering, College of Engineering, Jeonbuk National University

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.

      • 4,000원

      0개의 논문이 장바구니에 담겼습니다.