earticle

논문검색

휴대용 맥박산소포화도 측정기를 이용한 맨손톱과 셀프 인조손톱의 산소포화도 비교

원문정보

Comparison of oxygen saturation between no polish nails and self-artificial nails using portable pulse oximeter

김수민, 이경열

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

Purpose: This study aimed to compare the oxygen saturation measured at fingertips with unpolished and self-artificial nails (sticker, tip) using a portable pulse oximeter. Methods: The sample comprised 27 women who had used nail care services. Using a pulse oximeter, oxygen saturation was measured at fingertips with unpolished nails in a stable state. Self-fabricated sticker and tip nails were attached to the right and left hands, respectively, using five nail colors (black, blue, green, purple, and brown). Oxygen saturation was measured at fingertips with self-fabricated nails first in the usual manner and then measured again by rotating the oximeter by 90°. Values of oxygen saturation were compared and analyzed using the Friedman test and Wilcoxon signed-rank test with Bonferroni correction using IBM SPSS 26.0. Results: Oxygen saturation values were significantly different among colors for both sticker nails (p=.036) and tip nails (p=.000), with black self-artificial nails tending to demonstrate lower oxygen saturation. Moreover, oxygen saturation values obtained by measuring the usual way on unpolished nails was significantly different than those obtained by rotating the oximeter by 90° on thumb with sticker (p=.001) and tip nails (p=.001). Conclusion: The findings of this study suggest that oxygen saturation should be measured on other fingers except for thumb instead of removing the self-artificial nails in emergency situations. For patients with black self-artificial nails, is the findings recommend measuring oxygen saturation at other body parts such as toes, earlobes, forehead, etc.

목차

= Abstract =
Ⅰ. 서론
1. 연구의 필요성
2. 연구의 목적
Ⅱ. 연구방법
1. 연구설계
2. 연구대상 및 자료수집 방법
3. 연구절차
4. 연구도구
5. 분석방법
6. 축약형 단어
Ⅲ. 연구결과
1. 연구대상자
2. 맨손톱의 손가락 간에 산소포화도 차이
3. 맨손톱과 셀프 인조손톱 간에 산소포화도 차이
4. 셀프 인조손톱 색깔 간에 산소포화도 차이
5. 맨손톱에서 일반적인 방법과 셀프인조손톱에서 측정기를 90° 회전한방법 간에 산소포화도 차이
Ⅳ. 고찰
Ⅴ. 결론
ORCID ID
References

저자정보

  • 김수민 Su-Min Kim. 충청남도 안전체험관 소방교
  • 이경열 Kyoung-Youl Lee. 국립공주대학교 응급구조학과 교수

참고문헌