earticle

논문검색

Original Article

광중합 방식의 적층 가공으로 제작된 전치과 구치 단일 크라운의 적합도 평가

원문정보

Evaluation of fit of anterior and posterior single crowns manufactured by light-curing additive manufacturing

배은정, 이완선

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

Purpose: This study aimed to evaluate the fit of the anterior and posterior teeth printed using two light-curing three-dimensional (3D) printers. Methods: Anterior and posterior single crowns were designed using dental software and were printed using 2 types of 3D printers, liquid crystal display (LCD) and digital light processing (DLP) (n=40). After the printed crown was scanned again from inside and outside, the prepared teeth were evaluated using a 3D program. To compare the root mean square (RMS) results among groups (α=0.05), the one-way analysis of variance and Tukey‘s test were used. Results: No statistically significant difference was found between the mean RMS values of the anterior and posterior teeth (p>0.05). However, as a result of comparing the internal, external, and tooth shapes, the DLP group showed significantly low errors in the inner and outer surfaces than LCD group (p<0.05). Conclusion: In terms of clinical acceptance standard of 100 μm, the fit of the anterior and posterior teeth fabricated using LCD and DLP was clinically acceptable.

목차

Article Info
INTRODUCTION
MATERIALS AND METHODS
1. 지대치 시편 선정
2. 3D 프린팅
3. 3D 적합도 분석
4. 통계분석
RESULTS
DISCUSSION
CONCLUSIONS
FUNDING
ACKNOWLEDGEMENTS
CONFLICT OF INTEREST
ORCID
REFERENCES

저자정보

  • 배은정 Eun-Jeong Bae. 부천대학교 치기공과
  • 이완선 Wan-Sun Lee. 부천대학교 치기공과

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.

      • 4,000원

      0개의 논문이 장바구니에 담겼습니다.