치아 교정용 호선의 순측 변위량에 의한 자가결찰 세라믹 브라켓의 인장력 분석

원문정보

Tensile Force Analysis of Self-Ligating Ceramic Bracket by Labial Displacement of Orthodontic Arch Wire

주형진, 이승주, 김원태, 김옥환

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

As the demand for aesthetic orthodontic treatment increases, the use of self-ligating ceramic brackets is increasing. For stable treatment, there should be no fracture or deformation of the self-ligation ceramic bracket door. Therefore, considering the situation in which labial displacement of teeth occurs in the orthodontic treatment stage. For this study, a model of the mandibular anterior region of a ceramic self-ligating bracket with a passive sliding door mechanism was selected. The measured tensile force data was substituted into the simulation analysis conditions, and the tensile force, stress distribution, and deformation values were analyzed using the finite element method. Using this, it is able to use the design elements of the orthodontic bracket that should be considered as design inputs in the development stage.

목차

ABSTRACT
1. 서론
2. 연구방법
2.1 도어 인장강도 시험
2.2 유한요소 모델 제작
3. 결과
3.1 인장강도 시험 결과
3.2 구조해석 결과
4. 결론
References

저자정보

  • 주형진 Hyung-Jin Ju. Dept. of Mechatronics, Kongju National University
  • 이승주 Seung-Ju Lee. Dept. of Future Convergence Engineering, Kongju National University
  • 김원태 Won-Tae Kim. Pfof. in the divisiion of Future Convergence Engineering, Kongju National University
  • 김옥환 Ok-Hwan Kim. Professor, Dept. of Mechanical and Automotive Engineering, Kongju National University

참고문헌