선택적이온질량분석기를 활용한 대기오염 배출원 관리의 신뢰도 향상연구
초록
영어
This study was conducted to improve the reliability of atmospheric source management using scientific equipment. Recently, Selected Ion Flow Tube Mass Spectrometers (SIFT-MS), a real-time air quality measurement equipment, was introduced to efficiently manage air pollution emission sources. SIFT-MS is not a measurement method by air pollution process test standards. Consequently, simultaneous measurements were carried out with the existing method of gas chromatography (GC)-mass spectrometry (MS). The 2 ppm standard gas used to calibrate the equipment was diluted to 10, 50, 100, and 200 ppb for analysis. Simultaneous measurements of SIFT-MS and GC-MS showed that differences in measurement concentrations were within 1 ppb. Standard gas analysis shows that all eight VOCs showed superior linearity with a correlation coefficient of 0.99 or higher. It confirmed that measurements of SIFT-MS were reliable and that SIFTMS can efficiently manage sources of suspected high concentrations by measuring air pollutants in real time.
한국어
본 연구는 과학적 장비를 활용한 대기배출원 관리의 신뢰성을 향상 시키기 위해 진행하였다. 최근 효율적으로 대기배 출원을 관리하기 위해 실시간 대기질 측정 장비인 선택적이온질량분석기(Selected Ion Flow Tube Mass Spectrometers, SIFT-MS)이 도입되었다. SIFT-MS는 대기오염공정시험기준에 의한 측정방법이 아니다. 그러므로 기존 측정방법인 기체 크로마토그래피(Gas Chromatography, GC)-질량분석기(Mass Spectrometry, MS)와 동시측정을 진행하고 장비를 교정하 는데 사용되는 표준가스 2 ppm을 10, 50, 100, 200 ppb로 희석하여 분석을 진행하였다. SIFT-MS, GC-MS 동시측정 결 과 측정농도의 차이는 약 1 ppb 이내로 나타났고 표준가스 분석결과 8가지 VOCs 모두 상관계수 0.99 이상의 우수한 직 선성을 나타냈다. SIFT-MS의 측정값이 신뢰도가 있음을 확인하고 SIFT-MS를 활용하면 실시간으로 대기오염물질을 측 정해서 고농도 의심 배출원을 효율적으로 관리 할 수 있음을 확인했다.
목차
요약문
I. 서론
II. 연구방법
1. 실시간 대기질 측정장비
2. 측정장소
3. SIFT-MS와 열탈착 흡착 튜브를 이용한 VOCs 측정
4. Syft standard gas를 이용한 농도별 VOCs 측정
III. 결과 및 고찰
1. SIFT-MS, GC-MS 측정결과
2. SIFT-MS의 검량특성 평가
IV. 결론
References