Design of ferromagnetic shims for an HTS NMR magnet using sequential search method

초록

영어

This study deals with the ferromagnetic shims design based on the spherical harmonic coefficient reduction method. The design method using the sequential search method is an intuitive method and has the advantage of quickly reaching the optimal result. The study was conducted for a 400 MHz all-REBCO magnet, which had difficulty in shimming due to the problem of SCF (screening current induced field). The initial field homogeneity of the magnet was measured to be 233.76 ppm at 20 mm DSV (Diameter Spherical Volume). In order to improve the field homogeneity of the magnet, the ferromagnetic shim with a thickness of 1 mil to 11 mil was constructed by a design method in which sequential search algorithm was applied. As a result, the field homogeneity of the magnet could be significantly improved to 0.24 ppm at 20 mm DSV and 0.05 ppm at 10 mm DSV.

목차

Abstract
1. INTRODUCTION
2. FERROMAGNETIC SHIMS DESIGN
2.1. Spherical harmonics and field mapping
2.2. Ferromagnetic shimming
2.3. Ferro-shims design with sequential search method
3. CASE STUDY ON SHIMMING DESIGN FOR 400 MHz NMR MAGNET
3.1. Field mapping results of the HTS NMR magnet
3.2. Process of the ferromagnetic shims design
3.3. Design result and virtual mapping
4. CONCLUSION
ACKNOWLEDGMENT
REFERENCES

저자정보

  • Hongmin Yang Kunsan National University, Jeonbuk, Gunsan, Korea, Korea Basic Science Institute, Chungbuk, Cheongju, Korea
  • SangGap Lee Korea Basic Science Institute, Chungbuk, Cheongju, Korea
  • Minchul Ahn Kunsan National University, Jeonbuk, Gunsan, Korea

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.
      ※ 학술발표대회집, 워크숍 자료집 중 4페이지 이내 논문은 '요약'만 제공되는 경우가 있으니, 구매 전에 간행물명, 페이지 수 확인 부탁 드립니다.

      • 4,000원

      0개의 논문이 장바구니에 담겼습니다.