원문정보
Comparison of changes in ventilation volume according to fixation method of I-gel during cardiopulmonary resuscitation : a study using a simulated manikin
초록
영어
Purpose: The I-gel device is Korea’s most frequently used airway management method during pre-hospital cardiopulmonary resuscitation (CPR). This study aimed to compare changes in ventilation volume according to the fixation method with a simulated manikin. Methods: We placed I-gel into an advanced life support simulator and compared tape and band fixation conditions. CPR was performed according to the 2020 Korean CPR guidelines, using a mechanical chest compression device and an adult bag. The positional shift of I-gel and the ventilation volume of the simulated manikin were measured after performing CPR for 20 minutes. Five trials were carried out in each setting. Statistical analysis was carried out with SPSS 27.0. P < .05 was considered significant. Results: Positional shift after 20 minutes of CPR was as follows: tape fixation, 7.2 ± 0.2 mm; band fixation, no change, indicating a significant difference between the two groups (p=.003). The mean ventilation volume was tape fixation, 482.63 ± 30.84 mL; band fixation, 544.96 ± 22.98 mL, showing a significant difference (p=.002). Conclusion: When using the I-gel during pre-hospital CPR, using a band-type fixing device with elasticity rather than fixing the tape provides stable and appropriate ventilation by maintaining the fixed position.
목차
Ⅰ. 서론
1. 연구의 필요성
2. 연구의 목적
Ⅱ. 연구방법
1. 연구대상
2. 자료수집 및 연구도구
3. 분석방법
Ⅲ. 연구결과
1. 아이젤 고정 시 소요 시간과 심폐소생술 후 고정 위치 변동
2. 심폐소생술 시 구간별 환기량의 변화
Ⅳ. 고찰
Ⅴ. 결론
ORCID ID
References