영어

In order to assess the cell toxicity of 10 instruments made of polymers, the MTT assay which utilizes the L-929 cell was selected. Specimens were eluted at a temperature of 37°C for 24 hours at a rate of 4g per 20mL, RPMI 1640, and then was positively and negatively contrasted with a control test solution, in accordance with the Notification No. 2020-12 Protocols of Medical Apparatus Biological Safety from the Ministry of Drug and Food Safety. As a result of 24 hours of incubation in 37°C, 5% CO2 Incubator and assessment using an ELISA reader, the results of Intraoral camera indiciated a cellular viability of more than 70% at a 50% eluate. But, the Plastic impression tray, 3D printing tweezer, Impression disposable syringe, Dental floss holder, Hand implant scaler, Surgical retractor, Oral scanner tip, Dental mirror, and the Water pick tip all reported a cellular viability of more than 70% at a 100% eluate, which indicates that do not exhibit cytotoxicity, thus allowing it to be used in contact with the mucous membrane of the oral cavity.

한국어

최근 치과에서 사용되는 10종의 폴리머 기구에 대한 세포독성을 평가하기 위하여 L-929 세포를 이용한 MTT 시험을 시행하였다. 검체를 4g 당 20mL의 비율로 제조한 37℃의 RPMI 1640 용액에서 24시간 동안 용출한 후 식품 의약품안전처 고시 제2020-12호 의료기기 생물학적 안전에 관한 공통기준규격에 따라 검체 용출액과 공시험액, 음성 및 양성대조를 사용하여, 37℃, 5% CO2 Incubator에서 24시간 배양하여 ELISA reader로 판정한 결과, Intraoral camera는 용출물 농도 약 50%에서 약 70% 이상의 세포 생존율을 나타냈나 Plastic impression tray, 3D printing tweezer, Impression disposable syringe, Dental floss holder, Hand implant scaler, Surgical retractor, Oral scanner tip, Dental mirror, Water pick tip은 모두 용출물 농도 100%에서 70% 이상의 세포 생존율로 세포 독성을 나타내지 않아 구강점막에 직접 접촉하여 사용이 가능한 기구로 평가되었다.

목차

Abstract
요약
1. Introduction
2. Materials and Methods
2.1 Study Materials and Specimen
2.2 Preparation of Experimental Materials
2.3 Preparation of control
2.4 Experimental Apparatus
2.5 Suppliers
2.6 Experimental procedures
2.7 Evaluation of cytotoxicity
3. Results
3.1 Intraoral camera
3.2 Plastic impression tray
3.3 3D printing tweezer
3.4 Impression disposable syringe
3.5 Dental floss holder
3.6 Hand implant scaler
3.7 Surgical retractor
3.8 Oral scanner tip
3.9 Dental mirror
3.10 Water pick tip
4. Discussion
5. Conclusions
REFERENCES