Accuracy evaluation of dental model scanner according to occlusal attrition type
초록
영어
Purpose: The purpose of this study is to compare and analyze the accuracy of single crowns based on the type of occlusal surface. Methods: A single crown wax pattern was fabricated in three types of occlusal surface. The prepared wax pattern was replicated with silicone, and stone was injected to create a stone model. The prepared specimens were scanned using a model scanner. Scans were classified into three groups, and each scan was performed six times to analyze the trueness and precision of a single crown. In addition, only the occlusal surface area was analyzed for trueness and precision. Data were analyzed using the Kruskal–Wallis H test, a nonparametric test (α=0.05). Results: With regard to the trueness value of the occlusal scan area, the no occlusal tooth attrition (NA) group showed the largest error of 3.5 μm, and the complete occlusal tooth attrition (CA) group showed the lowest value of 3.1 μm. The NA group had the greatest precision, and the medium occlusal tooth attrition (MA) group and CA group showed a low precision value of 3.2 μm; the difference between the groups was statistically significant (α=0.05). In the color difference map, the CA group showed a lower error than the NA group. Conclusion: The occlusal surface with severe attrition had excellent accuracy, but the accuracy of the group without attrition was low. There were significant differences between groups, but clinically acceptable values were shown.