Ka 밴드에서 Power Meter 계측 명령어에 따른 측정 정확도와 소요시간에 대한 연구

원문정보

A Study on Measurement Accuracy and Required Time based on SCPI of Power Meter in Ka Band

조태종, 신석호

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

Measurement accuracy and required time is important to make ATE(Automatic test equipment) system in Ka band, and SCPI commands of power meter which is a representative RF test equipment are studied in this paper. Comparison data between FETCH and MEASURE which are SCPI commands are measured in 30 G ~ 31 GHz and -70 ~ +20 dBm using two power sensor. The data show that FETCH which is the fastest SCPI is able to get reliable data in linear interval above noise level. MEASURE which is the best accurate command takes longer time than FETCH, and the longest time is 13.2 seconds. These results offer that measurement accuracy and required time of the two SCPI for power meter and would be used as a guideline for efficient ATE system in Ka band.

한국어

측정 명령어에 따른 정확도와 소요시간은 효율적인 Ka 밴드용 자동 시험 장비(ATE)를 구축하는데 중요한 요소이 다. 이에 따라 본 논문에서는 RF 대표 계측기 Power Meter의 측정 명령어에 대해 연구하였다. 30 G ~ 31 GHz에서 Power Meter의 측정 명령어 FETCH와 MEASURE의 각 측정 레벨에 따른 정확도와 소요시간을 비교하였으며, -70 ~ +20 dBm 영역을 2가지 Power Sensor로 측정하였다. 측정 결과 신속하게 데이터를 불러오는 FETCH 명령어는 잡 음 레벨보다 높은 선형 구간에서 비교적 정확한 데이터를 얻을 수 있었다. 가장 정확한 데이터를 얻을 수 있는 MEASURE 명령어는 FETCH 대비 시간이 많이 소요 되었으며, 최대 13.2초가 소요 되었다. 이를 통해 30 G ~ 31 GHz 에서 Power Meter 측정 명령어의 정확도와 측정 소요시간을 확인할 수 있었고, 해당 연구 결과는 효율적인 Ka 밴드용 ATE를 구축하는데 기준을 제시할 수 있을 것으로 기대된다.

목차

요약
Abstract
Ⅰ. 서론
Ⅱ. 시험 구성 설명
Ⅲ. 측정 결과
1. N8487A (-30 ~ +20 dBm)
2. 8487D (-70 ~ -20 dBm)
3. N8487A와 8487D의 측정 결과 고찰
Ⅳ. 결론
References

저자정보

  • 조태종 Tae-Chong Cho. 정회원, 국방기술품질원
  • 신석호 Suk-Ho Shin. 정회원, 국방기술품질원

참고문헌

자료제공 : 네이버학술정보