국산 치과용 구강스캐너(e-scanner)와 모델스캐너의 정확도 비교

원문정보

Comparison of the accuracy of domestic dental intra-oral scanner(e-scanner) and model scanner

김부섭, 김정호

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

Purpose: The purpose of this study is to evaluate the discrepancy of scan process in dental intra oral scanner by comparing model scanner and anticipate possibility to introduce intra oral scan technique. Methods: 3D superimposition test was conducted to compare the scan discrepancy. The scanners used in this study are the e-oral scanner, the D750 model scanner, and the high precision CMM(3D Coordinate Measuring Machine). The standard of accuracy verification is ISO 5725-1; trueness and precision. Master model was manufactured by dental stone and scanned 5 times by intra oral, model scanner. Reference data was scanned 5 times by high accuracy CMM to evaluate the trueness. Results: Trueness of D750 scanner were 7.4 ㎛ 5.1 ㎛ 6.8 ㎛ at an abutment, an occluasal, a specific area. and trueness of e-scanner were 20.2 ㎛ 27.4 ㎛ 37.8 ㎛ at an abutment, an occluasal, a specific area. Precision of D750 scanner was 7.04 ㎛, e-scanner was 15.95 ㎛. Conclusion: When conducting in vitro test, The mean difference of trueness between e-scanner and D750 were 12.8 ㎛ at an abutment area, 22.3 ㎛ at an occlusal area, 31.0 ㎛ at a specific area and 8.91 ㎛ in precision. The scan discrepancies are within the range of clinical acceptance.

목차

[Abstract]
I. 서론
Ⅱ. 연구 방법
1. 실험 재료
2. 실험 방법
Ⅲ. 결과
1. 지대치 중첩 결과
2. 교합면 중첩 결과
3. 인접 영역 중첩 결과
4. 정밀도(precision) 분석 결과
Ⅳ. 고찰
Ⅴ. 결론
REFERENCES

저자정보

  • 김부섭 Busob Kim. 부산가톨릭대학교 치기공학과
  • 김정호 Jungho Kim. 부산가톨릭대학교 치기공학과

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.

      • 4,000원

      0개의 논문이 장바구니에 담겼습니다.