AEC를 적용한 복부 CT 검사 시 환자 자세와 피폭선량에 대한 융합 연구

원문정보

The convergence study on patient position and exposure dose in abdominal CT examination using AEC

이춘규, 오정섭, 최선욱, 김갑중, 유세종, 전민철

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

The purpose of this study was to evaluate the dose and image quality according to the rotation of the X-axis direction in the abdominal CT scan, and to find ways to reduce the exposure dose. The phantom was scanned by rotating in the X-axis direction at 0, 5, 10, and 15 degrees, respectively. The CTDIvol value, HU, noise, and signal-to-noise ratio were measured at each rotation. ANOVA analysis was performed using the SPSSWIN (ver 19.0) program. The radiation exposure dose was 5.44mGy, 5.70mGy, 5.98mGy and 6.38mGy at 0, 5, 10 and 15 degrees, respectively. HU, noise, and signal-to-noise ratio were not statistically significant. In the CT scan, if the patient is located in the isocenter of the gantry aperture and there is no rotation in the X-axis direction, the exposure dose is reduced.

한국어

본 연구는 복부 전산화단층촬영 검사에서 X-축 방향의 회전에 따른 선량, 영상의 화질을 평가하여 피폭선량의 감소 방안을 평가 하였다. 인체 모형 팬텀을 이용해 0도, 5도, 10도, 15도로 5도씩 X-축 방향으로 회전하여 scan을 하였다. 각각의 회전에서 CTDIvol 값, HU, 노이즈, 그리고 신호대잡음비를 측정하였다. SPSSWIN(ver 19.0) 프로그램을 사용하여 ANOVA 분석을 하였다. 회전에 따른 선량은 0도, 5도, 10도, 15도 회전에서 각각 5.44mGy, 5.70mGy, 5.98mGy, 6.38mGy 측정되었다. 회전에 따른 HU, 노이즈, 그리고 신호대잡음비는 통계적으로 유의한 차이가 없었다. 전산화단층촬영 검사에서 환자가 gantry aperture의 isocenter에 위치함과 동시에 X-축 방향으로 회전이 없다면 피폭 선량이 감소되어진다.

목차

요약
Abstract
1. 서론
2. 연구 방법
2.1 실험장비 및 기구
2.2 실험방법
3. 연구 결과
3.1 회전에 따른 선량 비교 분석
3.2 회전에 따른 영상 평가
4. 고찰
5. 결론
REFERENCES

저자정보

  • 이춘규 Chun-Kyu Lee. 충남대학교병원 영상의학과
  • 오정섭 Jeong-Sub Oh. 충남대학교병원 영상의학과
  • 최선욱 Seon-Wook Choi. 충남대학교병원 영상의학과
  • 김갑중 Gab-Jung Kim. 송호대학교 방사선학과 교수
  • 유세종 Se-Jong Yoo. 대전보건대학교 방사선학과 교수
  • 전민철 Min-Cheol Jeon. 대전보건대학교 방사선학과 교수

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.

      • 4,000원

      0개의 논문이 장바구니에 담겼습니다.