원문정보
초록
영어
The purpose of this study was to evaluate the dose and image quality according to the rotation of the X-axis direction in the abdominal CT scan, and to find ways to reduce the exposure dose. The phantom was scanned by rotating in the X-axis direction at 0, 5, 10, and 15 degrees, respectively. The CTDIvol value, HU, noise, and signal-to-noise ratio were measured at each rotation. ANOVA analysis was performed using the SPSSWIN (ver 19.0) program. The radiation exposure dose was 5.44mGy, 5.70mGy, 5.98mGy and 6.38mGy at 0, 5, 10 and 15 degrees, respectively. HU, noise, and signal-to-noise ratio were not statistically significant. In the CT scan, if the patient is located in the isocenter of the gantry aperture and there is no rotation in the X-axis direction, the exposure dose is reduced.
한국어
본 연구는 복부 전산화단층촬영 검사에서 X-축 방향의 회전에 따른 선량, 영상의 화질을 평가하여 피폭선량의 감소 방안을 평가 하였다. 인체 모형 팬텀을 이용해 0도, 5도, 10도, 15도로 5도씩 X-축 방향으로 회전하여 scan을 하였다. 각각의 회전에서 CTDIvol 값, HU, 노이즈, 그리고 신호대잡음비를 측정하였다. SPSSWIN(ver 19.0) 프로그램을 사용하여 ANOVA 분석을 하였다. 회전에 따른 선량은 0도, 5도, 10도, 15도 회전에서 각각 5.44mGy, 5.70mGy, 5.98mGy, 6.38mGy 측정되었다. 회전에 따른 HU, 노이즈, 그리고 신호대잡음비는 통계적으로 유의한 차이가 없었다. 전산화단층촬영 검사에서 환자가 gantry aperture의 isocenter에 위치함과 동시에 X-축 방향으로 회전이 없다면 피폭 선량이 감소되어진다.
목차
Abstract
1. 서론
2. 연구 방법
2.1 실험장비 및 기구
2.2 실험방법
3. 연구 결과
3.1 회전에 따른 선량 비교 분석
3.2 회전에 따른 영상 평가
4. 고찰
5. 결론
REFERENCES