earticle

논문검색

Research on the Dispersion Coefficient Measurement System based on CCD Imaging Automatic Alignment Technology

초록

영어

In order to measure the dispersion coefficient of glass and other transparent material rapidly and accurately, the paper studies the dispersion coefficient measurement system based on CCD imaging automatic alignment technology. On the basis of consideration of the air refractive index impacting on the measurement accuracy of dispersion coefficient, this paper rebuilds the mathematical model by the traditional dispersion coefficient measurement formula. Using the measuring angle technology of precision grating, the paper realizes the digital measurement of deflection angle of refracted ray. Sub-pixel edge detection algorithm is adopted to realize the automatic alignment of double line to single line, which has reduced artificial error of the dispersion coefficient in the measurement process. Finally, the paper combined the multichannel small collimator designed technology and the modern control technology to achieve the high precision measurement of refractive index. Demarcating the measurement system by known dispersion coefficient of optical glass, testing results and error of standard is less than ±2.3×10-6 absolutely.

목차

Abstract
 1. Introduction
 2. The New Dispersion Coefficient Measurement Model
 3. Dispersion Coefficient Measurement System Composition and Working Principle based on CCD Imaging Automatic Alignment
 3. CCD Imaging and Image Process Algorithm
 5. The Measurement Results
 6. Analysis of Measuring Precision
 7. Conclusion
 References

저자정보

  • Cao Miao Changchun University of Science and Technology, Changchun, 130022, China
  • Yao Qinghua Changchun University of Technology, Changchun, 130012, China

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 원문제공기관과의 협약기간이 종료되어 열람이 제한될 수 있습니다.

      0개의 논문이 장바구니에 담겼습니다.