earticle

논문검색

Research on Size Precision of Implant Fixtures among Four Different Types of Fastening Methods

초록

영어

This study carried out size dimensional test selecting regular size implant fixtures that have similar upper diameter and the longest implant fixtures among four different types of fastening methods processed surface with SLA. Size test result of implant fixture upper diameter of 4 kinds by each conclusion method, all test pieces showed small variation less than standard size ± 1%, which is the norm standard size in exam specification standards of dental implant Dimensional medical devices test at the Republic of Korea Food and Drug Administration.

목차

Abstract
 1. Introduction
 2. Research Methods
  2.1. Test Methods
  2.2. Analysis of Test Results
 3. Results
 4. Discussion
 5. Conclusions
 References

저자정보

  • Soo-chul Park Department of Dental Technology, Gimcheon University,
  • Chang-suk Kim Department of Dental Hygiene, Ulsan College 101 Bongsuro Donggu Ulsan, 682-715, Korea
  • In-ho Jeong Department of Dental Technology, Gimcheon University

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 원문제공기관과의 협약기간이 종료되어 열람이 제한될 수 있습니다.

      0개의 논문이 장바구니에 담겼습니다.