The purpose of this study is to compare the differences between Shear compression load using shear compression loads tester (858 Bionix, MTS, USA) with fixed screws in the force of 30 N·cm using in four different fastening methods with implanting(N=5) fixture an abutment. As a result of the experiment, the compression load of YI Implant which is the compression load of Internal octagon connection Implant , Non submerged type taking morse taper of 8° was the highest as 1208.20 N·cm, that of Internal hexagon connection Implant which is submerged type taking morse taper of 11° was the lowest as 617.60 N·cm. The relevance of the shear compression load test which was done by four different fastening methods with implanting fixture an abutment, showed a statistically significant difference (p <0.001).