earticle

논문검색

방사선치료계획 정도관리 방법에 따른 신뢰도 평가

원문정보

Reliability estimation about quality assurance method of radiotherapy planning

김정호, 김가중, 유세종, 김기진

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

According as radiation therapy technique develops, standardization of radiation therapy has been complicated by the plan QA(Quality Assurance). However, plan QA tools are two type, OADT (opposite accumulation dose tool) and 3DADT (3 dimensional accumulation dose tool). OADT is not applied to evaluation of beam path. Therefore tolerance error of beam path will establish measurement value at OADT. Plan is six beam path, five irradiation field at each beam path. And beam path error is 0 degree, 0.2 degree, 0.4 degree, 0.6 degree, 0.6 degree, 0.8 degree. Plan QA accomplishes at OADT, 3DADT. The more path error increases, the more plan QA error increases. Tolerance error of OADT path is 0.357 using tolerance error of conventional plan QA. Henceforth plan QA using OADT will include beam path error. In addition, It will increase reliability through precise and various plan technique.

목차

Abstract
 1. 서론
 2. 대상 및 방법
  2.1 대상
  2.2 방법
 3. 결과
  3.1 대향형 정도관리 평가
  3.2 3차원 정도관리 평가
  3.3 허용오차의 설정
 4. 결론 및 고찰
 5. References