earticle

논문검색

융합 의과학

사이버나이프에서 Xsight spine tracking system을 이용한 3D 표적위치보정의 유용성 평가

원문정보

Useful evaluation of 3D target location correction by using Xsight spine tracking system in CyberKnife

정영준, 김상현

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

The purpose of this study is to evaluate utility of rotating adjustment using Xsight spine tracking system in 3D DOF location adjusting method, to minimize error between 6D DOF and 3D DOF in fiducial tracking system. In this study, the result of 6D DOF target location error is 0.124 ± 0.058mm, using fiducial inside tumor 3D DOF 0.673 ± 0.142mm, outside tumor 1.126 ± 0.253mm, apply with Xsight spine tracking system 3D DOF 0.542 ± 0.103mm. As the experiment shows, it was demonstrated that rotating adjustment through Xsight spine tracking system is valuable in case of treatment in 3D DOF location error that makes increasing accuracy and dose distribution each approximately 48% and 3%. In accordance with result of this study is useful rotation.

한국어

사이버나이프치료에서 삽입된 금표지자의 삽입 개수 및 인식의 제한으로 인하여 3D DOF로 치료하는 경우 척추구조물 정렬을 가능하게 하는 Xsight spine tracking system으로 회전방향의 위치오차를 보정함으로서 표적위치오차를 개선하고, 치료방법의 대안을 제시하고자 한다. 실험결과 6D DOF에서 표적위치오차는 0.214 ± 0.058 ㎜, 종양내부에 삽입된 2개 금표지자를 이용한 3D DOF에서 0.673 ± 0.142 ㎜, 종양외부에 삽입된 2개 금표지자를 이용한 3D DOF에서 1.126 ± 0.253 ㎜, Xsight spine tracking system의 적용 한 3D DOF에서 0.542 ± 0.103 ㎜로 나타났다. 실험결과 척추 구조물로 회전방향에 대한 보정을 시행하였을 때 표적위치에 대한 정확성이 약 48% 향상되었다. 또한, 선량분포의 일치성도 약 3%가 향상되어 일치하는 것을 확인할 수 있었다. 따라서 척추구조물 정렬을 병용한 Xsight spine tracking system의 회전방향에 대한 보정은 유용한 것으로 평가 되었다.

목차

요약
 Abstract
 1. 서론
 2. 대상 및 방법
  2.1 Radiochromic 필름을 이용한 E2E test
  2.2 필름 교정
  2.3 선량분포 분석
 3. 결과
  3.1 Radiochromic 필름을 이용한 E2E test 결과
  3.2 필름 교정 결과
  3.3 선량분포 결과
 4. 고찰 및 결론
 REFERENCES

저자정보

  • 정영준 Young-Joon Jeong. 순천향대학교병원 방사선종양학과, 고려대학교 의학물리학과
  • 김상현 Sang-Hyun Kim. 서울대학교병원 영상의학과, 을지대학교 방사선학과

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.

      • 4,000원

      0개의 논문이 장바구니에 담겼습니다.