Determination of cobalt(III) ion with a perfluorinated sulfonated polymer-ethylenediamine (nafion-en) modified glassy carbon electrode is studied. It is based on the chemical reactivity of an immobilized layer, nafion-en, to yield complex [Co(en)3]3+. The reduction peak potential by differential pulse voltammetry (DPV) is observed at - 0.437±0.047 V (vs. Ag/ AgCl). The linear calibration curve is obtained in cobalt(III) ion concentration range 1.0×10-8~1.0× 10-3 M (5.893×10-12~ 5.893×10-5 g/mL).