earticle

논문검색

요추 측면 검사 시 공간선량 분포와 피폭선량

원문정보

Spatial dose distribution and exposure dose during lumbar lateral test

김창규

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

During the lateral x-ray testing of lumbar, in order to obtain the optimal image for diagnosis and to minimize the exposure dose, a glass dosimeter and spatial dose measuring meter was used to measure and evaluate the exposure dose and spatial dose distribution of each organs. The exposure dose of the organs have increased as they were closer to the X-ray tube and when the radiation field was completely opened, the exposure dose was increased. In addition, scattered rays have increased as the distance got closer to the subject and with the distance of more than 200cm, 95% of scattered rays was reduced. Such results can anticipate the exposure dose of patients during the lumbar x-ray test in the future and it can be proposed as a data for determining the testing methods and expected to be widely used as an important basic data for reducing the medical exposure dose.

한국어

요추 측면 X선 검사 시 피폭선량을 최소화 하고 진단에 최적의 영상을 얻기 위하여 유리선량계와 공간선 량 측정계를 이용하여 장기별 피폭선량과 공간선량분포를 측정 평가 하였다. X선관과 가까울수록 장기의 피폭선량 이 증가하였으며 조사야를 완전히 열었을 때 피폭선량이 증가되었다. 또한 피사체와 거리가 가까울수록 산란선이 증가하였으며 200 cm 이상 거리를 두면 95% 이상의 산란선이 감소되었다. 이 결과는 향후 요추 X선 검사 시 환자 의 피폭선량을 예측하고 검사 방법을 결정하는데 자료로 제시되어 의료피폭선량을 감소 하는데 중요한 기초자료로 많은 활용이 있을 것으로 기대된다.

목차

요약
 Abstract
 1. 서론
 2. 연구대상 및 방법
  2.1 재료 및 장비
  2.2 유리선량계 피폭선량 측정
  2.3 공간산란선량 측정
  2.4 통계 처리 및 분석
 3. 결과 및 고찰
  3.1 요추측면 검사 시 피폭선량
  3.2 요추 측면 검사 시 거리변화에 따른 공간선량 분포
  3.3 요추 측면 검사 시 측정 방향에 따른 공간 선량 분포
 4. 결론
 REFERENCES

저자정보

  • 김창규 Chang-Gyu Kim. 김천대학교 방사선학과

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.

      • 4,000원

      0개의 논문이 장바구니에 담겼습니다.