진단용 X선 장치에서 방사선출력의 재현성 평가 (IEC 60601-2-54 표준규격을 기반으로)

원문정보

Evaluation of the Reproducibility of Radiation Output from Diagnostic X-ray Equipment(Standards Based on IEC 60601-2-54)

한범희, 정홍량, 임청환, 김종일, 이상호, 한상현, 홍동희, 김창규, 유인규, 모은희

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

For five diagnostic X-ray generators (DR), four units turned out to be appropriate in tests on the reproducibility of radiation output suggested in the IEC 60601-2-54 standard, but in one unit of the X-ray equipment, an item measured in a combination of 50% of the highest tube voltage of the diagnostic X-ray equipment, the test setting of Group C with authorized output doses between 1 μGy and 5 μGy of mAs turned out to be inappropriate. As a result, the radiation dose to the IEC 60601-2-54 standard for quantification standards proposed by the radiation output from diagnostic X-ray imaging device reproducibility of performance management should be aware that an important evaluation factor.

한국어

총 5대를 대상으로 한 진단용X선발생장치(DR)는 IEC 60601-2-54 표준규격에서 제시한 방사선출력의 재현성 시험에서 4대가 적합으로 나타났으며, 다만 1대의 X선 장치에서 C Group의 시험 설정인 진단용X선장치의 최고 관전압의 50%와 관전류 시간 곱(mAs)이 1μGy에서 5μGy 사이의 인가출력량의 조합으로 측정한 항목에서 부적합으로 나타났다. 이에 따라 방사선량을 정량화하기 위한 표준으로 IEC 60601-2-54 표준규격에서 제시한 방사선출력의 재현성이 진단용 X선 영상장치 성능관리에 중요한 평가 요소임을 인식해야 한다.

목차

요약
 Abstract
 1. 서론
 2. 실험 및 방법
  2.1 실험기기
  2.2 실험방법
 3. 실험결과
 4. 고찰
 5. 결론
 ACKNOWLEDGMENTS
 REFERENCES

저자정보

  • 한범희 Beom-Hee Han. 서남대학교 방사선학과, 전북대학교 방사선과학기술학과
  • 정홍량 Hong-Ryang Jung. 한서대학교 보건의료학과
  • 임청환 Cheong-Hwan Lim. 한서대학교 보건의료학과
  • 김종일 Chong-Yeal Kim. 전북대학교 방사선과학기술학과
  • 이상호 Sang-Ho Lee. 서남대학교 방사선학과
  • 한상현 Sang-Hyun Han. 서남대학교 방사선학과
  • 홍동희 Dong-Hee Hong. 서남대학교 방사선학과, 한서대학교 보건의료학과
  • 김창규 Chang-Gyu Kim. 김천대학교 방사선학과
  • 유인규 In-Gyu You. 한서대학교 보건의료학과, 한림대학교병원 영상의학과
  • 모은희 Eun-Hee Mo. 전북대학교 방사선과학기술학과

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.

      • 4,000원

      0개의 논문이 장바구니에 담겼습니다.