earticle

논문검색

Measurement of thermal contact resistance at Cu-Cu interface

초록

영어

The thermal contact resistance (TCR) is one of the important components in the cryogenic systems. Especially, cryogenic measurement devices using a cryocooler can be affected by TCR because the systems have to consist of several metal components in contact with each other for heat transferring to the specimen without cryogen. Therefore, accurate measurement and understanding of TCR is necessary for the design of cryogenic measurement device using a cryocooler. The TCR occurs at the interface between metals and it can be affected by variable factors, such as roughness of metal surface, contact area and contact pressure. In this study, we designed TCR measurement system at various temperatures using a cryocooler as a heat sink and used steady state method to measure the TCR between metals. The copper is selected as a specimen in the experiment because it is widely used as a heat transfer medium in the cryogenic measurement devices. The TCR between Cu and Cu is measured for various temperatures and contact pressures. The effect of the interfacial materials on the TCR is also investigated.

목차

Abstract
 1. INTRODUCTION
 2. EXPERIMENTS
 3. RESULTS AND DISCUSSION
 4. CONCLUSION
 ACKNOWLEDGMENT
 EXAMPLE REFERENCES

저자정보

  • Myung Su Kim Korea Basic Science Institute, Daejeon, Korea
  • Yeon Suk Choi Korea Basic Science Institute, Daejeon, Korea

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 기관로그인 시 무료 이용이 가능합니다.