초록
영어
In this paper, we address the critical issues raised inmobile ECG monitoring and propose new solutions to resolve respective problem. A chest-belt type two-electrode wireless ECG measurement system and the specific android ECG monitoring application are presented. Furthermore, we also propose optimum filter-set for motion artifact removal in ECG recording. Average R-peak detection success rate is seen to be improved from 71.8% up to 97.4% after deploying the optimum filter-set coefficient. Besides that, we also suggest a simple and reliable arrhythmia ECG detection algorithm. The algorithm is proved with a promising accuracy of 95.4% in detecting arrhythmia ECG pattern. Lastly, a real time experiment is presented to demonstrate mobile ECG monitoring using a 10.1 inch Galaxy Tab.
목차
1. Introduction
2. Hardware Architecture
2.1. A New Analog Front-end Low Pass Input Buffer
2.2. ECG Signal Conditioning Circuit
2.3.Mobile ECG Monitoring using 10.1 inch Galaxy Tab
2.4. A New ECG Monitoring Experience for Android Tablet
3. Experiment and Results
3.1. Optimum Filter-set Coefficient
3.2. Arrhythmia ECG Detection
3.3 Real Time ECG Monitoring
4. Conclusions
Acknowledgements
References