치과용 백색광 스캐너를 이용한 impression scanning의 반복 측정에 대한 안정성 평가

원문정보

Evaluation on the repeatability of dental white light scanner-based digital impression

전진훈, 이경탁, 김혜영, 김지환, 김웅철

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

Purpose: The purpose of this study was to evaluate the repeatability of dental white light scanner. Methods: The impression(Zerosil, Dreve, Germany) were digitized in white light scanner(Identica, Medit, Korea) to create 3-dimensional surface-models. The distribution of the discrepancies between the number of points in the corresponding CRM models and the point clouds in the others were measured by a matching-software(Power- Inspect 2012, Delcam Plc, UK). The discriptive statistics were used for statistical analysis(SPSS 20.0). Results: The measurement of repeatablity showed very good reliability. The mean(SD) discrepancy value on the white light scanner digital models was 8.7(0.67) , based on SD and absolute mean values. Conclusion: These in vitro studies showed that repeatability of dental white light scanner is high reliability. These results can be confirmed in further clinical studies.