치과용 스캐너 전용 석고를 이용하여 제작된 3차원 디지털 모형의 정확도 평가

원문정보

An evaluation of validity of three dimensional digital model fabricated by dental scannable stone

김기백, 김수진, 김재홍, 김지환

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

Purpose: The purpose of this study was to evaluate the validity of digital models fabricated by dental scannable stone. Methods: Twenty same cases of stone models(maxillary full arch) were manufactured. Intercanine distance, intermolar distance, two dental arch lengths(right, left), two diagonal of dental arch lengths(right, left) were measured for comparison. Each of ten stone models were measured by digital vernier calipers and scanned by dental scanner. Ten digital models were measured by CAD program. The mean(SDs) values were compared by a Mann- Whitney U test( =0.05). Results: No statistically significant differences between the two groups were found at intermolar distance, dental arch length(right)(p>0.05). However, intercanine distance, dental arch length(left) and two diagonal of dental arch lengths(right, left) were statistically significant(p<0.05). Conclusion: Stone models fabricated by dental scannable stone showed larger than digital models.

목차

Abstract
 Ⅰ. 서론
 Ⅱ. 연구 방법
  1. 시편 제작
  2. 계측지점 선정
  3. 석고 모형의 계측지점 측정
  4. 디지털 모형의 계측지점 측정
  5. 통계 분석
 Ⅲ. 결과
 Ⅳ. 고찰
 Ⅴ. 결론
 REFERENCES