Three-dimensional evaluation on the repeatability and reproducibility of dental scanner-based digital models
초록
영어
Purpose: The aim of this study was to determine the repeatability and reproducibility of two dental scanners. Methods: The master die and the stone replicas(Kavo, Germany) were digitized in touch-probe scanner(Incise, Renishaw, UK), white light scanner(Identica, Medit, Korea) to create 3-dimensional surface-models. The number of points in the point clouds from each reading were calculated and used as the CAD reference model(CRM). Discrepancies between the points in the 3-dimensional surface models and the corresponding CRM were measured by a matching-software(Power-Inspect R2, Delcam Plc, UK). The t-student test for one samples were used for statistical analysis. Results: The reproducibility of both scanner was within 3 , based on mean value. The mean value between measurements made directly on the touch probe scanner digital models and those made on the white light scanner digital models was 2.20-2.90 , and was statistically significant(P<0.05). Conclusion: With respect to adequate data acquisition, the reproducibility of dental scanner differs. Threedimensional analysis can be applied to differential quality analysis of the manufacturing process as well as to evaluation of different analysis methods.