Mechanical Analysis of Hexagonal Porous Body for Porous Dental Implant
초록
영어
Purpose: The purpose of this study is a porous cube mechanical analysis for the dental implant. Methods: The porous cube with a side length of 10 was designed for dental implant. To choose proper design, porous hexagon with a side 10 which was drilled as a regular hexagon with diameter 0.8 , 1.0 , 1.2 and a side 0.4 , 0.5 , 0.6 each using Computer AUTO CAD(Autodesk, 2008). Each cube was carried out in the mechanical analysis. Results: The result of mechanical analysis was observed that the H0.8 was minimum stress 0.045068㎫, maximum stress 9.4565㎫ and minimum strain 0.00389 10-4㎫, maximum strain 0.816 10-4㎫, the H1.0 minimum stress 0.001147㎫, maximum stress 9.099㎫ and minimum strain 0.000099 10-4㎫, the maximum strain 0.784 10-4㎫, the H1.2 minimum stress 0.099393㎫, maximum stress 13.137㎫ and minimum strain 0.0112 10-4㎫, maximum strain 1.13 10-4㎫. Conclusion: The mechanical analysis of porous hexahedron was that H1.0 is the best result. It will be applicable to the porous implants.
목차
Ⅰ. 서론
Ⅱ. 연구 방법
1. 실험 재료
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
참고문헌