earticle

논문검색

Correction of Left Ventricle Strain Signals Estimated from Tagged MR Images

초록

영어

Strain measurement is a quantity used for assessing the regional function of the left ventricular (LV) of the heart. They are computed by tracking the motion of the non-invasive, virtual tags in the cardiac muscle with time. Tracking these tags gives information for each region of the cardiac muscle by quantifying its deformation during contraction (systolic period) and relaxation (diastolic period). However, these strain measurements suffer from inaccuracies caused by the degradation of the tags and the image quality. In this work, numerical simulations are used to investigate the factors contributing to the error in measurements. An empirical model for the estimated strain values is deduced and presented. In addition, a correction method for the measurement errors is proposed based on the empirically-deduced model. The method was validated on real data, and showed potential enhancement by reducing the errors in strain measurements.

목차

Abstract
 1. Introduction
 2. Theory and Methods
  2.1. Study of inherent causes of artifacts
  2.2. Strain Curve Modeling
  2.3. Weighted least square curve fitting
 3. Results
 4. Discussion
 5. Conclusion
 References

저자정보

  • Mina Khali Center for Informatics Science, Nile University, Cairo, Egypt
  • Ahmed Fahmy Center for Informatics Science, Nile University, Cairo, Egypt
  • Nael Osman Center for Informatics Science, Nile University, Cairo, Egypt, Johns Hopkins University, School of Medicine, Baltimore, Maryland, USA

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 원문제공기관과의 협약기간이 종료되어 열람이 제한될 수 있습니다.

      0개의 논문이 장바구니에 담겼습니다.