Implant에 사용되는 CAD/CAM 전용 zirconia abutment의 표면적과 디자인에 따른 유지력 차이

원문정보

Differences in retention of the reduction direction depending on

남택모, 김한곤, 김병식, 임시덕

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

The aim of this study is to make some basic materials to find retention force difference based on the total height of CAD/CAM zirconia abutment used for implant, retention force difference based on how to regulate the height of the abutment, retention force difference based on the size and retention force difference based on the design group.
The retention force was measured by being pulled at the speed of 1mm/min after being combined with zirconia block and abutment using Temp-BOND of Kerr. The experiment was done at the research lab of OSTEM in Busan by useing UNIVERSAL TESTING MACHINE on March 3rd, 2010. The final conclusion was reached as follows through analyzing the amessed results.
1. After analysing the total height and the retention force, p-value had minor difference by 0.01 statistically.
Namely, 3㎜, 4㎜, 5㎜ had the retention force difference and we could see retention force difference on 3㎜ and 5㎜ at the post test.
2. After analyzing how to regulate the height and retention force, p-value had minor difference by 0.000 statistically. Namely, 1㎜ and 2㎜ had the retention difference and we could see that 1㎜ and 2㎜ with the total height had retention difference.
3. After analyzing the retention force based on the size, p-value had minor retention force difference by 0.000 statistically. Namely, 7 different size had retention force difference and we could see the size 21.9㎜, 32.9㎜, 32.9㎜,
38.4㎜ , 48.9㎜ and 54.9㎜ had retention force difference.
4. After analyzing the retention force based the design, p-value had retention force difference by 0.000 statistically. Namely 9 different design group had retention difference and we could see that 9 design group with 5.6.7.8 design group and 9 design group with 1.2.3.4. design group had retention force.

목차

[Abstract]
 Ⅰ. 서론
 Ⅱ. 연구∙실험 대상 및 방법
  1. 실험 재료
  2. 실험 장비
  3. ziconia 시편 제작 방법
  4. abutment의 종류에 따른 유지력 실험 방법
 Ⅲ. 실험 결과
  1. abutment 표면적과 디자인(군)에 따른 유지력 결과
  2. abutment 전체 높이와 유지력의 비교
  3. abutment 조절 높이와 유지력의 비교
  4. abutment  표면적과 유지력의 비교
  5. abutment 디자인(군)에 따른 유지력의 비교
  6. abutment 유지력과 높이, 표면적, 디자인(군)의 상관관계
 Ⅳ. 결과 및 고찰
 Ⅴ. 결론
 참고문헌