Response Function of HPGe Detector using 23Na(p, γ)24Mg and 27Al(p, γ)28Si Reaction
초록
영어
In the present work, peak relative efficiency for the energy was obtained and response function was worked out. This study was carried out using the high resolution high efficiency HPGe detector(diameter 78.7 mm, length 86.5 mm) and NaI(Tl) detector for anti-compton. The anti-coincidence of the signals from the two detectors could be used to lessen the Compton effect signal; thus, the γ-ray energy resolution could be improved. The γ-ray spectrum was measured at 55° to the direction of the incident proton beam. Reaction spectrum was obtained from the 23Na(p, γ)24Mg reaction at Ep = 1424 keV and 27Al(p, γ)28Si reaction at Ep = 992 keV. To accelerate the incident proton which creates the (p, γ) capture reaction, the 3 MeV Pelletron accelerator at the Tokyo Institute of Technology was used. Response function was worked out by a noble technique. We worked out a response function from 1.2 to 9.4 MeV at intervals of 0.75 MeV.
한국어
본 연구에서는, 에너지에 따른 peak의 상대효율을 구하였으며, 검출기의 응답함수를 작성하였다. 이를 위해 고효율, 고분해능을 가진 HPGe 검출기(지름 78.7 mm, 길이 86.5 mm)를 이용하였으며 콤프턴 억제용으로 NaI 검출기를 사용 하였다. 감마선 스펙트럼은 23Na(p, γ)24Mg 및27Al(p, γ)28Si 반응을 이용하여 얻었으며, 이 때 입사 입자의 에너지는 각각 Ep = 1424 keV 및 Ep = 992 keV 이었다. 한편 스펙트럼 측정은 입사 빔 방향에 대해 55°에서 하였으며, 사용한 가속기는 일본 동경공업대학의 3 MeV Pelletron 가속기를 이용하였다. 검출기의 응답함수는 1.2 MeV에서 9.4 MeV까지 0.75 MeV 간격으로 작성하였다.