Evaluation of a Fabricated Charge Sensitive Amplifier for a Semiconductor Radiation Detector

원문정보

초록

영어

A CSA(Charge Sensitive Amplifier) was designed and fabricated for application in a radiation detection system based on a semiconductor detector such as Si, SiC, CdZnTe and etc.. A fabricated hybrid‐type CSA was evaluated by comparison with a commercially available CSA. A comparison was performed by using calculation of ENC (Equivalent Noise Charge) and by using energy resolutions of fabricated radiation detectors based on Si. In energy resolution comparison, a fabricated CSA showed almost the same performance compared with a commercial one. In this study, feasibility of a fabricated CSA was discussed.

목차

ABSTRACT
 INTRODUCTION
 EXPERIMENTS
 RESULTS AND DISCUSSION
 CONCLUSION
 ACKNOWLEDGEMENTS:
 REFERENCES

저자정보

  • HAN SOO KIM Korea Atomic Energy Research Institute
  • JANG HO HA Korea Atomic Energy Research Institute
  • SE HWAN PARK Korea Atomic Energy Research Institute
  • JAE HYUNG LEE Korea Atomic Energy Research Institute
  • CHEOL HO LEE Korea Atomic Energy Research Institute

참고문헌