earticle

논문검색

포스터 3 분 Speech : 좌장: 장재형 교수(연세대학교), 강택진 교수(동국대학교)

Parylene film as a linker layer of SPR biosensor

초록

영어

Surface Plasmon resonance (SPR) biosensor has been applied for the on-line monitoring of a target analyte in complex mixture by using specific interaction between antigens and antibodies. In this work, mainly for reusability of SPR biochip, parlyene thin film was applied as a linker layer on gold surface of the SPR biochip. Parylene is a polymer of p-xylene with good optical properties, chemical resistance from organic solvents and relatively high mechanical strength which are required for the linker layer of SPR biosensor. The SPR effect is known to be the occurrence of an evanescence field by the absorption of incident light at the range of 100nm from the sensor surface and we tested the feasibility of parylene layer of less than 100nm as a linker
layer by observation of resonance angle change according to refractive index. Parylene film can be easily coated and removed by oxygen plasma treatment so that SPR biochip can be used repeatedly. The applicability of the parylene film was demonstrated for the detection
of anti-CCP(Citrullinated Cyclic Peptide) antibodies in the sera from rheumatoid arthritis(RA) patients.

저자정보

  • Byoung-Jin JEON Dept. of Materials Science and Engineering, Yonsei University
  • Moo-Hwan KIM Femto Science Co.
  • Jae-Chul PYUN Dept. of Materials Science and Engineering, Yonsei University

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 원문제공기관과의 협약기간이 종료되어 열람이 제한될 수 있습니다.

      0개의 논문이 장바구니에 담겼습니다.