Finite element stress analysis on supporting bone by tripodal placement of implant fixture
초록
영어
This study was to propose the clear understanding for stress distribution of supporting bone by use of staggered buccal offset tripodal placement of fixtures of posterior 3 crown implant partial dentures. We realized posterior 3 crown implant fixed partial dentures through finite element modeling and analysed stress effect of implant arrangement location to supporting bone under external load using finite element method.
Method: To understand stress distribution of 3 crown implant fixed partial dentures which have 2 different arrangement by finite element analysis. In each model, for loading condition, we applied 45° oblique load to occlusal surface of crown and applied 100 N for 3 crown individually(total 300 N) for imitating possible oral loading condition. at this time, we calculated Von Mises stress distribution in supporting bone through finite element method.
Result: When apply 45° oblique load to in-line arrangement model, maximum stress result for 100 N for each 3 crown 47.566 . In tripodal placement, result for 1 buccal offset tripodal placement implant model was maximum distributed load 51.418 , so result was higher than in-line arrangement model.
Conclusion: In stress distribution result by placement of implant fixture, the most effective structure was in-line arrangement. The tripodal placement does not effective for stress distribution, gap cause more damage to supporting bone.