Accuracy Evaluation of CyberKnife SynchronyTMRespiratory Tracking SystemUsing Phantom
초록
영어
This study was conducted to evaluate the accuracy of CyberKnife SynchronyTMrespiratory tracking system which wasapplied to Stereotactic Radiosurgery (SRS) for moving tumors in chest and abdomen with breathing motion. For accurate evaluation, gold fiducial marks were implanted into a moving phantom. The moving phantom was a cube imbedding an acryl ball as a target. Theacryl ball was prescribed to 20 Gy at 70% of isodose curve in a virtual treatment and radiochromic films were inserted into the acrylball for dose verification and tracking accuracy evaluation. The evaluation of position tracking consists of two parts: fiducial marktracking in a stationary phantom and SynchronyTMrespiratory tracking in a moving phantom. Each measurement was done in threedirections and was repeated to 5 times. Range of position error was 0.1957 mm to 0.6520 mm in the stationary phantom and 0.4405 mmto 0.7665 mm in the moving phantom. Average position error was 0.3926 mm and 0.5673 mm in the stationary phantom and themoving phantom respectively. This study evaluates the accuracy of CyberKnife SynchronyTMRespiratory tracking system, and confirmsthe usefulness when it’s used for Stereotactic Radiosurgery of body organs.
목차
1. 서론
2. 재료와 방법
2.1 사이버나이프 방사선수술 장치의 소개
2.2 Synchrony TM 호흡추적 장치
2.3 Phantom을 이용한 정확성평가
3. 결과
3.1 고정된 Phantom을 이용한 금침추적 방법의 표적 정확성 평가결과
3.2 움직임 Phantom을 이용한 Synchrony TM 호흡추적방법의 표적 정확성평가 결과
4. 고찰
5. 결론
참고문헌
abstract