The present study was intended to examine a basic scheme to determine the biochemical oxygen demand (BOD) of livestock wastewater by means of six individual dissolved oxygen (DO) sensors and its multi-measurable meter. Maximal point of the first order time derivative of the DO difference between DO distribution of sterilized livestock wastewater and that of non-sterilized livestock wastewater, was considered as the oxygen uptake rate (OUR) of microorganisms in livestock wastewater, as determined to be 0.00074 mg O2/ℓ․sec. The present study showed that there was a fair linear relationship (97.72%) between maximal OUR and BOD values of livestock wastewater, the latter being determined by classical Winkler azide method. It was thus concluded that the present multi-biosensor system might be applicable to an on-line system for measurement of BOD of livestock wastewater.
한국어
용존산소 센서 및 시스템에서 손실되는 산소량을 보정하기 위해 멸균처리 기법을 도입하였다. 멸균처리 시료의 DO 거동과 비멸균처리 시료의 DO 거동차 (△DO)는 시스템의 영향과 무관하게 폐수 자체의 순수한 미생물만이 소모하는 산소변화량을 구하였다. 용존산소 센서로 축산폐수의 시간에 따른 DO거동 (-d△DO/dt)을 측정하여 구한 폐수자체의 미생물에 의한 산소소모속도는 0.00074mg O2/ℓ․ sec이었다. 축산폐수의 원액을 희석하여 다양한 BOD값을 갖도록 시료를 제조한 후 DO meter로 측정하여 구한 DO 변화량은 동일한 시료를 5일 BOD 측정방법인 Winkler Azide화 변법으로 구한 BOD값과 선형구간 (30〜60분)에서 97.72%의 높은 상관 선형성을 보였다. 따라서 본 연구의 multi-biosensor 시스템은 축산폐수의 BOD를 짧은 시간에 정확하게 측정할 수 있는 가능성을 제시하였다.
목차
서론
재료 및 방법
시료 채취 및 BOD 비교 측정
시스템 구성 및 제어
멸균기법을 이용한 시스템 보정
산소 소모 속도 (Oxygen Uptake Rate; OUR)
축산폐수 미생물의 농도
측정센서의 보정
측정 방법
결과 및 고찰
BOD 측정법의 비교 결과
측정 센서 시스템의 보정
멸균기법을 이용한 시스템의 보정
한 지점의 축산폐수 DO변화량의 재현성
멸균기법을 이용한 축산폐수의 BOD 연속 측정