earticle

논문검색

Quartz Crystal Microbalance 시스템을 이용한 저밀도 지질단백질측정용 면역센서의 개발

원문정보

Development of Quartz Crystal Microbalance-Based Immunosensor for the Determination of Low-Density Lipoprotein

김상현, 윤현철, 김학성

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

Immunosensor for the determination of LDL(Low-Density Lipoprotein), a good indicator for the diagnosis of atherosclerosis and hypercholesterolemia, was developed by using quartz crystal microbalance(QCM). The immunosensor consists of flow-through cell, oscillating circuit, oscilloscope, and frequency counter. FIA(Flow Injection Analysis) was applied to the QCM system for the measurement of LDL in liquid phase. Antibody showing binding affinity against LDL was immobilized on the gold electrode of a quartz crystal by covalent coupling via polyethylenimine / glutaredehyde. LDL was injected and bound to the antibody immobilized on the QCM immunosensor. The response of the immunosensor (F0 - F1) was found to be proportional to the LDL concentration from 200ug /ml to 800ug /ml. Operational conditions for the operation of immunosensor were also investigated in terms of sensitivity and non-specific binding.

목차

서론
 재료 및 실험
  재료
  QCM 시스템
  Flow-through Cell
  항체의 고정화 방법
  측정방법
 결과 및 고찰
  응답 특성 곡선
  고정화방법들 간의 민감도(sensitivity) 비교
  비특미적 결합에 의한 효과 비교
  항원-항체 결합에 의한 효과 비교
  항체의 농도에 따른 고정화 효율의 최적화
  ELISA 에 의한 QCM 시스템내의 항원-항체 결합 역가 확인
 요약
 감사
 참고문헌

저자정보

  • 김상현 Sang-Hyun Kim. 한국과학기술원 생물과학과
  • 윤현철 Hyun C. Yoon. 한국과학기술원 생물과학과
  • 김학성 Hak-Sung Kim. 한국과학기술원 생물과학과

참고문헌

자료제공 : 네이버학술정보

    함께 이용한 논문

      ※ 원문제공기관과의 협약기간이 종료되어 열람이 제한될 수 있습니다.