축방향과 반경방향흐름 크로마토그래피의 단백질 분리속도와 분리능에 대한 비교

원문정보

Comparison of Axial and Radial Flow Chromatography on Protein Separation Speed and Resolution

김윤하, 이은규

피인용수 : 0(자료제공 : 네이버학술정보)

초록

영어

The relationship between pressure drop and liquid flow rate, for an axial and a radial flow chromatographic column packed with compressible porous media was theoretically analyzed using modified Kozeny-Carman equation. The results were compared with experimental observations obtained using compressible DEAE-agarose as a model medium. At 2-9 psi range studied, the theoretical derivation accounting for 'gel compression' effect predicted simple Langmuirian type response of volumetric flow rate to changes in pressure drop. On the other hand, the experimental response was more or less sigmoidal. At the same pressure drop, radial column showed 2-3 times higher flow rates than those of axial column both theoretically and experimentally. Using r-HBsAg crude extract, protein resolution effects between the two types of columns at different flow rates were compared side-by-side. It turned out that, though general chromatographic behavior was very similar, axial column was somewhat superior in terms of r-HBsAg recovery yield and specificity. However, the number of theoretical plates analysis indicated the protein resolution effects were comparable.

목차

ABSTRACT
 서론
  크로마토그래피에서 유속과 압력강하에 대한 이론적 고찰
  형상비(aspect ratio)에 따른 유속과 압력강하에 대한 이론적 고찰
 재료 및 방법
 결과 및 고찰
  압력강하가 유속에 미치는 영향
  DEAE 크로마토그래피에 의한 r-HBsAg 정제성능비교
 결론
 감사
 참고문헌