원문정보
Separation Characteristics of Oligodeoxynucleotides by High-Performance Membrane Chromatography
초록
영어
Oligodeoxynucleotides (ODNs) were separated by high-performance membrane chromatography (HPMC), a combined system of chromatography and membrane. The separation mechanism involved anion-exchange, and the stationary phase was cation CIM (Convective Interaction Media) DEAE disk (16×3 mm). Two types of mobile phase were used,
buffer A (20mM Tris-HCI, pH 7.4) and buffer B (buffer A + 1M NaCl). As the amount of NaCl dissolved in buffer linearly increased, the retention time shortened, which enabled a gradient elution mode. Based on the number of theoretical plates and resolution observed, the optimum mobile phase and operating condition (Buffer A/Buffer B=50/50 - 20/80 vol%, gradient time 2 min) were experimentally determined. In this experimental condition, ODNs were separated within 2 min at a mobile phase flow rate of 6 ml/min.
한국어
크로마토그래피와 막의 혼합형태인 HPMC를 사용하여 ODNs를 분리하였다. 분리 메카니즘은 음이온 교환작용이며 정지상은 양이온의 CIM DEAE disk (16×3 mm)이었다. 이 동상은 buffer A (20 mM Tris-HCI, pH 7.4)와 buffer B(buffer A + 1M NaCl)를 사용하였으며 ODNs를 분리하기 위해서 구배용매 조성법을 사용하였다. Buffer내 NaCl의 양이 선형적으로 증가할수록 체류 시간이 빨라졌기 때문에 구배용매 조성법으로 ODNs의 체류인자를 조절하여 분리하였다. 각 이동상의 조성에 따른 이론단수와 분리도를 비교한 결과 최적의 이동상 조성 (Buffer A/Buffer B=50/50 - 20/80 vol%, gradient time 2 min)을 실험적으로 얻었고 6ml/min의 이동상 유속에서 2분내에 ODNs를 분리할 수 있었다.
목차
서론
재료 및 방법
실험재료
실험기기
실험방법
결과 및 고찰
요약
REFERENCES