원문정보
Modification of Retention Factor of Mononucleotides by Compositions of Buffers and Methanol in RP-HPLC
초록
영어
Due to the advantage of RP-HPLC with a variety of compositions of mobile phases, experiments on water-soluble charged species were examined. The samples were mononucleotides (5'-CMP, 5'-UMP, 5'-GMP, 5'-IMP, 5'-AMP), and the buffers used were sodium phosphate monobasic and acetic acid. The concentrations of buffers ranged from 0.01 to 10 mM, while that of the methanol, an additive to the mobile phase was 5 to 20 vol.%. To predict the retention
factor of a sample in terms of its methanol composition (M, vol.%) and buffer(CB, mM), the following nonlinear equation is suggested,
k =a + b C B( 1+ c C B ) M d
where a, b, c, and d were experimentally determined constants. The regression coefficients were above 0.96, and the agreement between experimental and calculated retention factors were relatively good.
한국어
Buffer의 특성에 맞는 체류인자에 관한 모델식을 얻기 위해서는 buffer와 양이온의 농도에 관한 관련식을 실험으로 결정해야 한다. 또한 model 식에서 비이온과 음이온의 체류인자를 나타내는 k0와 k-의 값이 일정한 경향을 나타내고 있기 때문에 buffer의 농도와 modifier의 농도를 동시에 고려한 model 식이 가능하게 된다. 본 연구에서 제안한 모델식 (식(7))을 이용하여 임의의 buffer의 농도에 따른 시료의 체류시간을 예측할 수 있다. 이러한 모델식을 이용하여 buffer를 이용한 RP-HPLC의 분석 및 분리조건을 예측할 수 있고 시행오차적인 방법보다 빠른 시간 내에 최적의 분석 및 분리조건을 얻을 수 있다.
목차
서론
이론
재료 및 방법
결과 및 고찰
Buffer 농도의 영향
메탄올 함량의 영향
실험식
요약
감사
사용기호
REFERENCES